Continued advances in techniques for intraocular lens (IOL) power calculations and measurements of axial lengths have led to more accurate IOL implantation during cataract surgery. Furthermore, improvements in surgical procedures, such as small-incision phacoemulsification and foldable IOL insertion, have minimized the occurrence of surgically induced astigmatism (SIA). These developments allow for better refractive correction, close to the targeted spherical equivalent. Cataract surgery is now regarded as a refractive surgery. However, many cataract patients have a significant degree of pre-existing corneal astigmatism. In fact, it has been reported that 19% of cataract patients have a corneal astigmatism above 1.5 diopters (D) and 2% have astigmatism greater than 4 D. 1 Despite the use of several surgical interventions for correcting astigmatism such as arcuate keratotomy, limbal relaxing incision, and laser in situ keratomileusis (LASIK), there are several drawbacks to these methods: risk of unintended corneal perforation, low predictability of remaining astigmatism, long-term refractive instability, and low time, effort, or cost effectiveness. 2-7 In contrast, toric IOL implantation is known to be a safe, predictable, and simple technique for correcting astigmatism. 8 Although early reports on toric IOL have demonstrated effective, immediate astigmatism correction, the subsequent appearance of IOL rotation has resulted in a reduction of this benefit. Indeed, it is known that a 30° rotation of the IOL in the capsular bag after implantation can increase the astigmatism compared to the pre-operative value. 9-12
Recent studies have reported good rotational stability of AcrySof toric IOL (Alcon Laboratories, Inc., Fort Worth, TX, USA). 13-16 However, the relationship of anterior capsule contraction and IOL rotation is not known. This paper seeks to evaluate the alignment of AcrySof toric IOL over time, as well as the relationship of anterior capsule contraction and IOL rotation.
From March 2009 to December 2009, 19 eyes of consecutive 19 patients underwent microcoaxial cataract surgery and AcrySof toric IOL implantation by a single surgeon. Informed consent was obtained from all patients after explanation of possible surgical consequences and the need for an additional exam for the evaluation of rotational stability. Informed consent was obtained from all patients, and the study was performed according to the Helsinki Declaration.
Before surgery, all patients had a complete ophthalmic examination, including visual acuity (VA) (LogMAR), tonometry, autorefraction, slit lamp examination, fundoscopy, specular microscopy, and corneal topography (ORB scan II ® ; Bausch and Lomb, Greenville, SC, USA). Patients with a senile cataract and a regular corneal astigmatism greater than 1.00 D prior to surgery were included. Among these patients, those who had a corneal disorder affecting VA or corneal astigmatism, with history of glaucoma or retinal disorder, or with a previous history of intraocular or refractive surgery were excluded.
The spherical IOL power and axial length were measured using the A-scan (Axis II ® ; Quantel Medical, Clermont-Ferrand, France) and corneal astigmatism using the manual keratometry (DM-4 ® ; Topcon Corp., Tokyo, Japan), respectively. Spherical IOL power was calculated using the SRK/T formula (A-constant 118.4). Cylindrical IOL power and axis were determined by a toric IOL calculator program (available at: http://www.acrysoftoriccalculator.com), which uses the patient’s data, including the spherical IOL power, manual keratometry, location of incision and surgically induced astigmatism (SIA), to calculate the IOL power and axis.
Limbal reference markings were made using the toric reference corneal marker (AE-2793S, ASICO) with the patient seated upright on the slit-lamp and operating table right before the operation. The target axis was marked using a Mandez ring and a Nuijts toric axis marker (AE-2740, ASICO) aligned with the pre-placed reference marks during the operation. One experienced surgeon performed all cataract operations under topical anesthesia. Continuous curvilinear capsulorhexis was performed under viscoelastic. Phacoemulsification was performed through a 2.2 mm temporal clear corneal incision using the Infinity Vision System (Alcon Laboratories, Inc.) and OZil Torsional Handpiece (Alcon Laboratories, Inc.). The toric IOL was implanted in the capsular bag using the Monarch injector system (Alcon Laboratories, Inc.). After irrigation and aspiration, the IOL axis was aligned with the limbal axis marks. The final incision size was 2.3 mm to 2.4 mm in all cases and no sutures were applied to the incision site. No intraocular complications, such as posterior capsule rupture, occurred.
One week, one month and three months after surgery VA was measured and automated refractometer, slit-lamp examination, and manual keratometry were performed. In all patients, the pupil was dilated to check the reference dots on the optic and slit-lamp retroillumination of the anterior segment was obtained using digital photographs at each follow-up visit. It was not possible to use any anatomical features to assist in checking that the horizontal was always in the same plane. However, because the camera position was fixed, and subject’s head was relatively stable with the use of chin and head rests, it was felt there would not be significant deviation from the horizontal between images. The images were analyzed using the image-editing tools of Adobe Photoshop CS2 software. An independent observer determined the axis of IOL and the area of the capsulorhexis aperture as described below.
1. The axis of the IOL was determined using a theoretical line that connected the identifiable reference dots that were located on two opposite ends of the optic (Fig. 1A). The angle between this theoretical line and the horizontal was compared to the pre-operative, intended angle on the first day after operation. From the second follow-up appointment, the angle was compared with the value of the previously measured angle.
2. The size of the opening of the anterior capsule was determined using the Magnetic Lasso Tool (Adobe photoshop), and the number of pixels within the selected area was computed.
3. For standardization, the pixels of the area of the anterior capsule opening were divided by the pixels of an imagined circle, which was drawn based on the diameter determined by the distance of the two innermost reference dots on the optic (Fig. 1B). We tried to determine if this standardized value, which represents the amount of anterior capsule contraction, correlated with the degree of misalignment of the IOL axis.
4. For the evaluation of the relationship between the anterior capsule contraction and the direction of IOL rotation, we divided the anterior capsule opening area into four quadrants based on the IOL axis. The quadrant number was assigned based on the location of the haptics and optic of IOL (Fig. 2A, 2D). A standardized value was calculated for each quadrant. The change of this value and of the direction of the IOL rotation between one week and one month after operation were analyzed.
Statistical analysis was performed using SPSS for Windows, version 17.0 (SPSS Inc., Chicago, IL, USA). The numeric variables were summarized using the mean and standard deviation. Between-group comparisons were performed using the Wilcoxon signed-rank test. Friedman test was used to analyze the correlation between the rotation of intraocular lenses and the lapse of time after operation. All
Nineteen eyes were included in this study. Eyes for which reference dots of the IOL optic could not be identified due to insufficient mydriasis or poor photo quality, were excluded from the analysis.
Among 19 eyes for which IOL rotation was identifiable at three months post-surgery, six eyes had underwent implantation with SN60T3, nine eyes with SN60T4, and four eyes with SN60T5. The amount of surgical induced astigmatism was 0.22 (0 to 0.75) D using the Cravy vector analysis. 18 Although the mean corneal astigmatism was not statistically different after the operation compared to before the procedure, the mean refractive astigmatism and visual acuity improved significantly (Table 1). The mean degree of toric IOL axis misalignment was 2.18 (± 2.02) degrees at the three-month follow-up.
Table 1 . Patient demographics, visual acuity, corneal and refractive astigmatism before and 3 months after operation
Variables | Preoperative | Postoperative | |
---|---|---|---|
Eyes (n) | 19 | ||
Age (years) | 62.21(±13.13) | ||
UCVA (LogMAR) | 0.72 (±0.36) | 0.10 (±0.12) | <0.001 |
BCVA (LogMAR) | 0.56 (±0.39) | 0.04 (±0.07) | <0.001 |
Corneal astigmatism (D) | 1.86 (±0.73) | 1.86 (±0.86) | 0.871 |
Refractive astigmatism (D) | 1.69 (±0.80) | 0.63 (±0.48) | 0.03 |
* Wilcoxon signed ranks test.
During follow-up, the number of eyes for which the direction and degree of IOL rotation could be determined and compared at one week, one month and three month after surgery was 16 (Table 2). The largest misalignment of the IOL axis was seen at the one day post-operative follow-up. The degree of misalignment decreased at each follow-up evaluation. Counterclockwise rotation was more frequent (10/16 eyes, 63%) at the first follow-up, while clockwise rotation was more frequent (11/16 eyes, 69%) at one week post-surgery. The frequency of each rotational direction was nearly equal at the later follow-up evaluations.
Table 2 . Misalignment and direction of the AcrySof toric intraocular lenses rotation with the lapse of time after operation
7 days | 1 month | 3 months | ||
---|---|---|---|---|
Degree of changes (n = 16) | 1.13 (±1.71) | 1.92 (±2.01) | 2.18 (±2.02) | < 0.001 |
Direction and amount of misalignment (degrees) | ||||
Clockwise (n = 8) | 1.26 (±2.27) | 2.14 (±2.13) | 2.45 (±2.22) | 0.023 |
Counterclockwise (n = 8) | 0.99 (±1.05) | 1.70 (±2.00) | 1.91 (±1.91) | 0.007 |
* Friedman test.
The degree of axis rotation recorded between the first postoperative day and three months following surgery did not correlate with the area of the anterior capsule opening, measured at post-operative day one, nor with the difference in the area of the anterior capsule opening between one-day and three-months post-surgery (
It is well known that AcrySof toric IOL implantation is an easy and effective tool for correcting corneal astigmatism, with good postoperative rotational stability. 13-17 Recent studies of this novel IOL have reported that the mean degrees of its angle of rotation are from 3.35 to 3.63. 13-15
The degree of IOL axis misalignment was largest one day post-surgery; this observation suggests that the cause of IOL rotation is strongly influenced by some peri-operative factors. These factors can be classified into 2 categories: failure to achieve an accurate alignment along the intended axis under the microscope during surgery or immediate rotation of IOL, despite an accurate insertion of the IOL.
AcrySof Toric IOL is an open-loop, modified single-piece, hydrophobic acrylic polymer material with L-shaped haptics. This IOL is designed to rotate clockwise during surgery, and adheres to the capsule after expansion. Thus, the surgeon first inserts the IOL at an axis that is minimally deviated, counterclockwise, from the intended axis. After removal of viscoelastic materials, the IOL is rotated, clockwise, to align with the intended axis as determined by the protocol. During this process, capsule adhesion of the IOL can occur before the final alignment and can result in a misaligned axis. Additionally, we experienced some error associated with the corneal reference marks. Marking color was diminished or, in some patients, spread during the operation, and this may also result in an inaccurate IOL insertion. One report found that a large mark size made it difficult to ensure accuracy, resulting in a misalignment of ±3 degrees of the intended axis for 95% of the markings. 19 Additional novel tools or methods are needed for more accurate marking.
The additional rotations that were sequentially measured after the first postoperative day were relatively minimal. These delayed rotations seem to be affected by various factors such as the size of the capsular bag, which is associated with the axial length, and the thickness of the IOL, although there was no statistical correlation between IOL rotation and axial length or IOL power in this study. In addition to these factors, IOL tilting and decentration, which was not observed in the current study, or progressive contraction of the capsule, may contribute to the gradual rotation for several months after an operation. Subcapsular fibrosis can cause contraction, which can result in the compression of the supporting area of the inserted IOL.
Significant rotation was also detected between the first and the seventh day follow-up evaluations. During this post-operative period, the direction was mainly clockwise. The mechanisms promoting this directionality are unclear but it may be due to the properties of the IOL to rotate clockwise. Considering the amount of rotation, this seems to be affected by a dynamic factor such as shrinkage of the capsular bag. Haptic compression or expansion may induce a torque on the optic causing it to rotate independently of the haptic. 20 Compression will cause clockwise rotation provided a moment is generated about the center of the IOL. Expansion would, in contrast, produce anticlockwise rotation. Although the capsular bag shrinkage was not seen between one- and seven-days post-surgery, as determined by slit-lamp examination, in this study, Strenn et al. 21 reported that the capsular bag diameter was reduced between one-day and one-week post-surgery, as determined by measuring the capsular tension ring. Therefore, this suggests rapid shrinkage of the capsular bag following implantation. This can also be accompanied by IOL rotation.
Gradual capsular contraction can lead to a delayed IOL rotation because the capsule tissue is attached to the IOL. Although, the total anterior capsule opening size did not have a statistical correlation with the degree of IOL rotation in this study, we could predict that IOL rotation was attributed to the capsule contraction. Because the quadrant analysis of the area of the anterior capsule opening showed a comparatively constant result. There was a tendency to counterclockwise rotation when contraction was dominant in haptic part (quadrant II or IV) of IOL and clockwise rotation when dominant in the non-haptic part (quadrant I or III). The cause of this trend is uncertain but one hypothesis is that since the haptics have a shape that extends in the counterclockwise direction from the IOL optic, when compressed, the entire IOL can rotate to the same direction. A clinical study has shown that a reduction in circumference of the capsular bag is universal, implying that compression of the haptic is more likely than expansion. 21 The counterclockwise rotation observed in our study is therefore likely to represent rotation of the haptic within the bag rather than that caused by haptic expansion. In contrast, when the compression is towards the center of the optic portion of IOL, the entire lens rotates clockwise due to the inherent properties of the IOL.
Several methods have been proposed to prevent continuous curvilinear capsulorhexis (CCC) contraction. Careful anterior capsular polishing and removal of the anterior subcapsular epithelial cells are important. An effective hydrodissection also helps to make it easier to remove the lens substance, ensuring a more complete removal of the cortex and associated cells. Large CCC (> 5.5 mm in diameter) may also be beneficial; small CCC have been found to correlate with lens capsular contraction. 22,23 Although the degree of IOL rotation decreased over time, there were a number of eyes that displayed a small amount of IOL rotation up until the three-month follow-up evaluation. Thus, additional rotation may occur over a longer time period and further observation is needed.
The accuracy of measuring IOL rotation can be questioned because of possible errors arising from any head tilt during photography or imprecise framing of images. That is the main limitation of this study. However, to minimize this error and to achieve objectivity during the determination of the angle of IOL misalignment and the amount of capsulorhexis aperture area, anterior segment photographs were taken by one experienced examiner with a same magnifying power (10×). Magnifying power compensation of the computerized images was done by an independent observer using our specific standardization protocol which was mentioned above. The use of digital imaging to assess rotation has been shown to be an accurate, visual method of evaluating IOL axis alignment that may reduce the errors of measuring IOL rotation with standard photography. 24 Both this study and the previous smaller-scale assessment of AcrySof SA60AT used a similar methodology for assessing rotation digitally. 25 Our results compare well with those in other studies of the same toric IOL design. The rotational stability of the AcrySof toric IOL in our study compares well with findings in theU.S. FoodandDrugAdministration (FDA) AcrySof Toric IOL clinical trial. The mean rotation in the FDA study was less than 4.0 degrees.
In conclusion, we have shown that the direction and the degree of AcrySof toric IOL rotational changes during the post-operative period. The most severe misalignment was observed one day after surgery and care should be taken to minimize the peri-operative risk factors that influence this early IOL rotation. There was a tendency to counterclockwise rotation when contraction was dominant in haptic (closed loop) part of IOL and clockwise rotation when dominant in the non-haptic (open loop) part.
The English in this document has been checked by International Science Editing, editing company. For a certificate, see: www.internationalscienceediting.com
The authors have no conflict of interest to disclose.
The present research was conducted by the research fund of Dankook University in 2018.
Kim JH, Cho KJ. Rotational stability of AcrySof toric intraocular lens over time: influence of capsulorhexis contraction. Med Laser 2020;9:44-50. https://doi.org/10.25289/ML.2020.9.1.44